CDW-60 impact test low-temperature instrumentDetailed product description:Column tabs
Product Introduction:
The CDW-60 impact test low-temperature instrument is made of 1Cr18Wi9Ti stainless steel plate for the inner chamber workbench and outer shell. This machine is mainly used for cooling impact specimens during low-temperature impact testing. Adopting a brand new compressor and a cascade cooling method, the operation is simple, safe, fast, reliable, and automatically maintains a constant temperature. When cooling, only power is needed without dry ice and liquid nitrogen. It has fast cooling speed, large volume, and high temperature control accuracy, making it an ideal supporting equipment for low-temperature impact testing.

2、 Technical standards

1. Technical parameters:
Temperature range: -60 ℃∽+30 ℃
Temperature fluctuation: ≤± 0.5 ℃
Temperature control instrument: digital display resolution: 0.1 ℃
Temperature sensor: PT100 platinum resistor
Temperature control element: thyristor
Heater: 1000W stainless steel heating tube
Vertical and horizontal mixing pumps: 6LITIES/MIN
Compressor: Operating current 2.9A
Refrigerant: R502F13
Opening size: 120 × 120 × 80mm (depth × width × height)
Dimensions: 105 × 530 × 86mm (depth × width × height)
Capacity: 10LIT1/CS
Digital display timer with a resolution of 1 second

2. Working conditions of CDW-60 impact test low-temperature instrument:
Power supply: 220V 50HZ 3KW
Medium used in the tank: ethanol or other freezing liquid
Environmental temperature: 5-32 ℃
Environmental humidity: ≤ 85%
The usage environment should be well ventilated, free from strong vibrations, dust, and corrosive gases