| measurement function |
| Measurement Mode |
|
Frequency A, Frequency B, Period A, RPM A, Totalize A. Frequency A/Ao, Frequency A-Ao, Pulse Width A | |
| Frequency A or B (optional) |
| Range: |
Frequency A: 0.1 Hz... 160 MHz |
|
Frequency B: 70 MHz... 1.3 GHz (Option PM 9608B) |
| pattern |
Countdown frequency calculation |
| Displayed LSD: |
2.5 x 10-7 x frequency/measurement time | |
| Frequency A/Ao |
|
Measure frequency A. The measured frequency is divided by a constant Ao before display. The resolution of the displayed ratio is determined by the measurement of frequency A. When powered on, Ao is set to 1 (default value). | |
| Frequency A-Ao |
|
Measure frequency A. Subtract the value of constant Ao from the measured frequency before displaying. The resolution of the displayed differential is determined by the measurement of frequency A. When powered on, Ao is set to 0 (default value). | |
| RPM A |
|
Measure frequency A. Multiply the measured frequency by 60 and display the result as revolutions per second (RPM) on the screen. |
| Range: |
6 RPM...720 x 106 RPM | |
| Cycle A |
| Range: |
8 ns...2 x 108s |
| pattern |
Single cycle measurement (SINGLE) or cycle average measurement (measurement time of 0.2, 1, or 10 seconds). |
| Displayed LSD: |
Single cycle measurement: |
|
(Time<100s): 100ns |
|
(Time>100s): 5 x cycles/109s |
|
Cycle average measurement: 2.5 x 10-7 x cycles/measurement time | |
| Accumulate A |
|
The calculation of the event is controlled by the START/STOP button. Accumulate the number of consecutive start stop cycles. The RESET button will turn off the gate signal and reset the frequency counter to zero. |
| Range: |
0... 1 x 1015 , There are units indicating k or M (thousands or trillions). If it exceeds the range, the result will be discarded. |
| Frequency range: |
Sine wave: 10 Hz... 16 MHz |
|
Pulse wave: 0 Hz... 16 MHz |
|
Pulse resolution: 80 ns |
| Displayed LSD: |
1 count (count<109) |
|
5 x Count/109 (Count ³ 109) | |
| Width A |
|
Measure the forward pulse width. The measurement time option is invalid (usually SINGLE measurement). |
| Range: |
100 ns...2 x 108s |
| Displayed LSD: |
(Time<100s): 100ns |
|
(Time ³ 100s): 5 x width/109s |
|
(Only when the duty cycle of the signal is 0.5, a trigger may occur at 50% of the amplitude) | |