HomeYour location:Homeinstrumentation and metersElectrical instrumentsfrequency meter → DH1001 Rubidium Atomic Frequency Standard
Column tabs 
Product name: DH1001 Rubidium Atomic Frequency Standard
specification: DH1001 Rubidium Atomic Frequency Standard
  Zoom Zoom in on the image
Category: instrumentation and meters -- frequency meter
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-139101609058

Email 91way@163.com
Wechat 13901609058(Wechat)


DH1001 Rubidium Atomic Frequency StandardDetailed product description:Column tabs
The DH1001 rubidium atomic frequency standard is a highly stable frequency source that utilizes the ground state hyperfine energy level transition spectral lines of rubidium isotope atoms to control the frequency of voltage controlled crystal oscillators. It is a passive atomic frequency standard. This instrument is suitable for time and frequency measurement, radio navigation and positioning, tracking of missiles and satellites, astronomical research, geological exploration, network synchronization of digital communication, precision timing and timing, etc.

1. Frequency accuracy:
a) Frequency calibration: The frequency accuracy at the factory should be better than 1 × 10-10.
b) Frequency adjustment range: The frequency adjustment range is greater than 2 × 10-9.
2. Frequency drift rate: The daily frequency drift rate is better than 2 × 10-12/d.
3. Frequency stability:

Sampling time τ 1.0 seconds 10.0 seconds 100.0 seconds
Stability σ y (τ) ≤ 1 × 10-11 ≤ 3 × 10-12 ≤ 1 × 10-12

4. Frequency startup characteristics: At 25 ℃, the lock time is ≤ 5 minutes; After 10 minutes of startup, the frequency startup characteristic is ≤ 2 × 10-10.
5. Frequency temperature characteristics: In an environment of -10 to+50 ℃, the frequency variation is ≤ 3 × 10-10.
6. The influence of geomagnetic field on frequency: The influence of geomagnetic field on frequency is ≤ 2 × 10-11.
7. Output frequency: 10MHz, 5MHz, 1MHz.
8. Output amplitude (sine wave): The effective voltage value on a 50 Ω load is 0.8V to 1.2V.
9. Harmonic distortion: ≤ -40dB.

DH1001 Rubidium Atomic Frequency Standard
Product Category
All product information on this website is translated by Baidu machine translation, which may not be completely accurate and is only for reference