As the latest generation of universal, multi-purpose, and multi range impedance testing instrument, AS2853A high-frequency Q meter has a testing frequency upper limit of 160MHz, which is currently the highest in China. The AS2853A high-frequency Q-meter adopts multiple leading technologies:
Dual scan technology - dual scan and automatic tuning search function for testing frequency and tuning capacitance.
Dual test element input - both the test frequency and tuning capacitance value can be entered through digital buttons.
Dual digital tuning - digital frequency tuning, digital capacitor tuning.
Automated measurement technology - automatically measure the Q value, resonance point frequency, and capacitance of the test piece.
Full parameter LCD display - digital display of main capacitance, inductance, Q value, signal source frequency, and resonant pointer.
DDS digital direct synthesis signal source - ensuring high fidelity of the signal source, high precision of frequency, and high stability of amplitude.
Computer automatic correction technology and test circuit optimization - minimize residual inductance in the test circuit, completely eliminating the confusion of correcting Q reading values at different frequencies.
AS2853A digital high-frequency Q-meter (100KHz -160MHz) main technical characteristics
| Q-value measurement range | 2~1023, range division: 30, 100, 300, 1000, automatic or manual shifting |
| Inherent error | ≤ 5% ± 2% of the full-scale value (200kHz~10MHz), ≤ 6% ± 2% of the full-scale value (10MHz~160MHz) |
| working error | ≤ 7% ± 2% of the full-scale value (200kHz~10MHz), ≤ 8% ± 2% of the full-scale value (10MHz~160MHz) |
| Inductance measurement range | 4.5nH ~ 140mH |
| Direct measurement range of capacitance | 1 ~ 200pF |
| Main capacitor adjustment range | 18 ~ 220pF |
| Accuracy of main capacitor adjustment | Below 100pF ± 1pF; Above 100pF ± 1% |
| Signal source frequency coverage range | 100kHz ~ 160MHz |
| Frequency segmentation (virtual) | 100 ~ 999.999kHz , 1 ~ 9.99999MHz,10 ~ 99.9999MHz , 100 ~ 160MHz |
Frequency indication error | 3 × 10-5 ± 1 word |