HomeYour location:Homeinstrumentation and metersElectrical instrumentsbridge → ZC2818A High frequency precision LCR digital bridge
Column tabs 
Product name: ZC2818A High frequency precision LCR digital bridge
specification: ZC2818A High frequency precision LCR digital bridge
  Zoom Zoom in on the image
Category: instrumentation and meters -- bridge
Column tabs 
Price: factory price
Brand: Shanghai
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-139101609058

Email 91way@163.com
Wechat 13901609058(Wechat)


ZC2818A High frequency precision LCR digital bridgeDetailed product description:Column tabs

ZC2818A high-frequency precision LCR digital bridge is an impedance measuring instrument with high measurement accuracy, wide testing range, and 6-digit testing resolution. It has a testing frequency of up to 300kHz, can be programmed in 0.01Hz steps, and has a signal level of 10mV~2V. It can meet the requirements of production line quality assurance, incoming inspection, and laboratory high-precision measurement. The HANDLER and RS232C interfaces provided by the instrument provide conditions for the instrument to be used in component automatic sorting systems, communicate with computers, and record testing processes.
Due to the absence of ideal inductors, capacitors, and resistors, in order to accurately evaluate the actual performance of an actual component in a certain frequency band, ZC2818A, with its signal generation capability of 10mHz resolution, can accurately analyze the frequency response of component parameters in any frequency band, especially for precise measurement of the resonant impedance of low-frequency ceramic resonators with a wide impedance range.

Performance characteristics:
·240 × 128 dot matrix large graphic LCD display
·Test frequency ranging from 20Hz to 300kHz, programmed in 0.01Hz steps at will
·10mVrms~2.0Vrms test level, programmed in 0.01V steps at will
·30 Ω, 100 Ω optional signal source output impedance
·Open circuit, short circuit, load correction function
·Built in comparator, 10 level sorting and counting functions
·RS232C, Handler interface

General indicators:


Working temperature and humidity
0℃~40℃, ≤90%RH
Power requirements
voltage
198V~242V
frequency
47.5Hz~63Hz
power consumption
≤50VA
volume
320mm×140mm×400mm
weight
About 3.5kg


Measurement function
test parameters
Z 、C、L、X、B、R、G、D、Q、θ
Basic accuracy
0.05%
equivalent circuit
Series connection and parallel connection
Mathematical functions
Absolute value deviation, percentage deviation
Range method
Automatic, hold
Trigger method
Internal, manual, external, bus
test speed
Fast: 32ms
Medium speed: 90ms
Slow speed: 650ms
Correction function
Open circuit, short circuit, load correction
Test end configuration
5 ends
average frequency
1-255
display mode
Direct reading ΔABS、Δ%、V/I

(Voltage/current monitoring), file number counting
monitor
240 × 128 dot matrix graphic LCD display with 6-digit reading resolution
Measurement signal
Measure signal frequency
20Hz-300kHz, Can be set arbitrarily in 0.01Hz steps
Measure signal level
0.01Vrms~2.00Vrms, Can be set arbitrarily in 0.01V steps
Signal output impedance
Choose between 30 Ω and 100 Ω
Measurement display range
|Z|、R、X
0.00001Ω~99.9999MΩ
C
0.00001pF~999.999mF
L
0.00001μH~9.99999kH
G、B
0.00001μS~999.999S
D
0.00001~9.99999
Q
0.00001~99999.9
θ(DEG)
-179.999°~179.999°
θ(RAD)
-3.14159~3.14159
Δ%
-999.999%~999.999%
Comparator and interface
comparator
Grade 11, Grade 9 is qualified, Grade 1 is unqualified. There is also one additional file attached
interface
RS232C, Handler, GPIB (optional)
ZC2818A High frequency precision LCR digital bridge