EN_1 digital display meter (EN_2)Detailed product description:

The WY2851D digital Q meter (50MHz) is a high-frequency impedance measurement instrument based on the principle of series resonance, which measures the high-frequency effective inductance, Q value, capacitance value, and ∑, tg δ of inductors, capacitors, insulation materials, etc The WY2851D digital Q meter has been digitized on the basis of the original WY2851Q meter, and the mechanical structure has been improved in terms of long service life.
Features:
*Our company's innovative automatic Q-value reading technology greatly improves the accuracy and speed of Q-measurement
*As low as 20nH residual inductance, ensuring minimal error in directly reading Q values at high frequencies
*Specially designed LCD screen with menu style display of multiple parameters, Q value, test frequency, tuning status, etc
*Q-value range automatic/manual range control
*Digital Q-value presetting can improve the reliability and speed of batch testing
*The VCO adjusts the test frequency, and its smooth continuous frequency variation enables convenient measurement of Q value for Δ f.
Main technical performance:
1. Operating frequency range:
50kHz~50MHz, Four digit effective digital display, divided into six frequency bands
2. Q-value measurement range:
10-1000, divided into three levels: 10-100, 30-300, 100-1000
3. Inherent error of Q value:
± 5% ± 3% full scale value
4. Effective inductance measurement range:
0.1μH~100mH。
5. Inductance measurement error:
≤ 5% ± 0.02 μ H
6. Tuning capacitor characteristics
6.1 Adjustable capacitance range:
40pf~500pf。
6.2 Accuracy:
± 1% or 1PF
6.3 Fine tuning capacitance:
± 3pf adjustable, ± 0.2pf resolution
6.4 Residual inductance value:
About 20nH
7. Q preset function:
Preset range: Q: 10-1000 is acceptable. After the tested component reaches the preset value, there will be a "GO" display and a beep sound prompt. Unqualified parts will display "NO GO"
8. Dimensions/Weight:
415 × 180 × 170 (mm), 7Kg.