HomeYour location:Homeinstrumentation and metersElectrical instrumentsQ table → EN_1 digital display meter (EN_2)
Column tabs 
Product name: EN_1 digital display meter (EN_2)
specification: EN_1 digital display meter (EN_2)
  Zoom Zoom in on the image
Category: instrumentation and meters -- Q Table
Column tabs 
Price: factory price
Brand: 无仪
Place of Origin: China
Column tabs 
Available Quantity: batch
delivery cycle: Spot goods (or inquire by telephone)
tabs
  Shanghai Beiyuan Industry and Trade Co.,Ltd
+86-21-66770508
+86-139101609058

Email 91way@163.com
Wechat 13901609058(Wechat)


EN_1 digital display meter (EN_2)Detailed product description:Column tabs
The WY2853D digital Q meter (50kHz-150MHz) is a high-frequency impedance measurement instrument based on the principle of series resonance, which measures the effective inductance, Q value, capacitance value, and ε, tan δ of inductors, capacitors, insulation materials, etc. at high frequencies. The WY2853D digital Q meter has been intelligently digitized based on the original WY2853 Q meter, expanding the measurement range and improving the long service life of the mechanical structure. This Q meter follows the national metrological verification regulations: "JJF1073-2000 High Frequency Q Meter Calibration Specification".

One characteristic:
Our company's innovative automatic Q-value reading technology greatly improves the accuracy and speed of Q-measurement.
DPLL synthesis generates test signals, with digital settings and continuous adjustment of signal frequency.
◎ Tune the residual inductance of the circuit to 8nH to ensure minimal error in directly reading the Q value at high frequencies.
Specially designed LCD screen with menu style display of multiple parameters: Q value, test frequency, tuning status, etc.
◎ Q-value range automatic/manual range control.
Digital Q-value preset can improve the reliability and speed of batch testing.
◎ Independent 50kHz~150MHz signal source output port, so this machine is also a 150MHz synthetic signal source, with frequency accuracy of ± 50ppm and stability of ± 10ppm.

Main technical specifications of WY2853D digital Q meter (50kHz-150MHz):
2.1 Operating frequency range: 50kHz~150MHz, five digit effective digital display.
2.2 Q-value measurement range: 1-1000; Divided into four levels: 1-30, 10-100, 30-300, 100-1000.
2.3 Intrinsic error of Q value: ± 5% ± 3% of full-scale value.
2.4 Effective inductance measurement range: 0.02 μ H to 200mH.
2.5 Inductance measurement error: ≤ 3% ± 0.01 μ H
2.6 Tunable capacitor characteristics:
2.6.1 Adjustable capacitance range: 13pF to 220 pF.
2.6.2 Accuracy: ± 1% or 0.5pF.
2.6.3 Capacitance measurement range: 1pF to 200pF.
2.6.4 Capacitance measurement error: ± 1% ± 0.5pF.
2.6.5 Residual inductance value: about 8nH.
2.7 Q preset function:
2.7.1Q preset range: 5-999 is acceptable. After the tested component reaches the preset value, there will be a "GO" display and a beep sound prompt. Unqualified parts will display "NO GO".
2.8 Dimensions and weight: 415 × 180 × 170 (mm), 7kg.